Portable Emissometer

Portable Emissometer

Devices and Services AE1 RD1 emissometer
CDevices and Services AE1 RD1 emissometer

For fast measurements of IR emissivity we have a Devices and Services AE1 RD1 emissometer.

Operation

It warms up time in 30minutes after which each measurement takes 90 seconds or less.

At the start of the measurement session the instrument is calibrated using a high (0.88) and low (0.05) reference sample.

To avoid inaccuracy from instrument drift the reference samples should be checked frequently.

If the sample is transparent in the IR range it is possible to obtain both emissivity and transmittance by measure using a dark and light background backing.

Results

The data produced by this instrument is a single number of the hemispherical IR emissivity, i.e. the value that is used in most thermal calculations.

Examples of Use

The AE1 RD1 has been used to measure the emissivity of a wide range of aluminum frame materials.

Emissivity of low-E coatings have been measured in agreement with FTIR inter-laboratory comparison.